Corneal Topography/Computer-Assisted Corneal Topography/Photokeratoscopy - CAM 90305
Description:
Computer-assisted corneal topography (also called photokeratoscopy or videokeratography) provides a quantitative measure of corneal curvature. Measurement of corneal topography is being evaluated to aid the diagnosis of and follow-up for corneal disorders such as keratoconus, difficult contact lens fits, and pre- and postoperative assessment of the cornea, most commonly after refractive surgery.
For individuals who have disorders of corneal topography who receive computer-assisted corneal topography/photokeratoscopy, the evidence includes only a few studies. Relevant outcomes are test accuracy, other test performance measures, and functional outcomes. With the exception of refractive surgery, a service not discussed herein, no studies have shown clinical benefit (e.g., a change in treatment decisions) based on a quantitative evaluation of corneal topography. In addition, a large prospective series found no advantage with use of different computer-assisted corneal topography methods over manual corneal keratometry. Computer-assisted corneal topography lacks evidence from appropriately constructed clinical trials that could confirm whether it improves outcomes. The evidence is insufficient to determine the effects of the technology on health outcomes.
Background
DETECTION AND MONITORING DISEASES OF THE CORNEA
Corneal topography describes measurements of the curvature of the cornea. An evaluation of corneal topography is necessary for the accurate diagnosis and follow-up of certain corneal disorders, such as keratoconus, difficult contact lens fits, and pre- and postoperative assessment of the cornea, most commonly after refractive surgery.
Assessing corneal topography is a part of the standard ophthalmologic examination of some patients.1,2 Corneal topography can be evaluated and determined in multiple ways. Computer-assisted corneal topography has been used for early identification and quantitative documentation of the progression of keratoconic corneas, and evidence is sufficient to indicate that computer-assisted topographic mapping can detect and monitor disease.
Various techniques and instruments are available to measure corneal topography: keratometer, keratoscope, and computer-assisted photokeratoscopy.
The keratometer (also referred to as an ophthalmometer), the most commonly used instrument, projects an illuminated image onto a central area in the cornea. By measuring the distance between a pair of reflected points in both of the cornea’s 2 principal meridians, the keratometer can estimate the radius of curvature of 2 meridians. Limitations of this technique include the fact that the keratometer can only estimate the corneal curvature over a small percentage of its surface and that estimates are based on the frequently incorrect assumption that the cornea is spherical.
The keratoscope reflects a series of concentric circular rings off the anterior corneal surface. Visual inspection of the shape and spacing of the concentric rings provides a qualitative assessment of topography.
A photokeratoscope is a keratoscope equipped with a camera that can provide a permanent record of the corneal topography. Computer-assisted photokeratoscopy is an alternative to keratometry or keratoscopy for measuring corneal curvature. This technique uses sophisticated image analysis programs to provide quantitative corneal topographic data. Early computer-based programs were combined with keratoscopy to create graphic displays and high-resolution, color-coded maps of the corneal surface. Newer technologies measure both curvature and shape, enabling quantitative assessment of corneal depth, elevation, and power.
Regulatory Status
A number of devices have been cleared for marketing by the U.S. Food and Drug Administration (FDA) through the 510(k) process. In 1999, the Orbscan® (manufactured by Orbtek, distributed by Bausch and Lomb) was cleared by the FDA. The second-generation Orbscan II is a hybrid system that uses both projective (slit scanning) and reflective (Placido) methods. The Pentacam® (Oculus) is 1 of a number of rotating Scheimpflug imaging systems produced in Germany. In 2005, the Pentacam HR was released with a newly designed high-resolution camera and improved optics. FDA product code: MXK.
Table 1. Corneal Topography Devices Clearing by the U.S. Food and Drug Administration
Device |
Manufacturer |
Date Cleared |
510.k No. |
Indication |
Galilei G6 Lens Professional | SIS AG, SURGICAL INSTRUMENT SYSTEMS | 07/25/2019 | K182659 | To scan, map and display the geometry of the anterior segment of the eye |
VX130 Ophthalmic Diagnostic Device | LUNEAU SAS | 4/24/2017 | K162067 | To scan, map and display the geometry of the anterior segment of the eye |
Pentacam AXL | OCULUS OPTIKGERATE GMBH | 1/20/2016 | K152311 | To scan, map and display the geometry of the anterior segment of the eye |
ARGOS | SANTEC CORPORATION | 05/16/201910/2/2015 | K191051K150754 | To scan, map and display the geometry of the anterior segment of the eye |
ALLEGRO OCULYZER | WAVELIGHT AG | 7/20/2007 | K071183 | To scan, map and display the geometry of the anterior segment of the eye |
HEIDELBERG ENGINEERING SLITLAMP-OCT (SL-OCT) | HEIDELBERG ENGINEERING | 1/13/2006 | K052935 | To scan, map and display the geometry of the anterior segment of the eye |
CM 3910 ROTATING DOUBLE SCHEIMPFLUG CAMERA | SIS LTD. SURGICAL INSTRUMENT SYSTEMS | 9/28/2005 | K051940 | To scan, map and display the geometry of the anterior segment of the eye |
PATHFINDER | MASSIE RESEARCH LABORATORIES INC. | 9/2/2004 | K031788 | To scan, map and display the geometry of the anterior segment of the eye |
NGDI (NEXT GENERATION DIAGNOSTIC INSTRUMENT) | BAUSCH & LOMB | 7/23/2004 | K040913 | To scan, map and display the geometry of the anterior segment of the eye |
PENTACAM SCHEIMPFLUG CAMERA | OCULUS OPTIKGERATE GMBH | 9/16/2003 | K030719 | To scan, map and display the geometry of the anterior segment of the eye |
ANTERIOR EYE-SEGMENT ANALYSIS SYSTEM | NIDEK INC. | 8/6/1999 | K991284 | To scan, map and display the geometry of the anterior segment of the eye |
ORBSCAN | TECHNOLAS PERFECT VISION GMBH | 3/5/1999 | K984443 | To scan, map and display the geometry of the anterior segment of the eye |
Related Policies
90314 Implantation of Intrastromal Corneal Ring Segments
90328 Corneal Collagen Cross-linking
Policy
Non-computer-assisted corneal topography is considered part of the evaluation/and management services of general ophthalmological services (CPT codes 92002 – 92014), and therefore this service should not be billed separately. There is no separate CPT code for this type of corneal topography.
Computer-assisted corneal topography is considered NOT MEDICALLY NECESSARY to detect or monitor diseases of the cornea.
Policy Guidelines
Please see the Codes table for details.
Benefit Application
BlueCard/National Account Issues
State or federal mandates (e.g., FEP) may dictate that all U.S. Food and Drug Administration (FDA)-approved devices, drugs, or biologics may not be considered investigational, and thus these devices may be assessed only on the basis of their medical necessity.
Some of these services may be provided as part of care that is not generally covered under health insurance contracts, such as:
- Contact lens fitting
- Refractive surgery
Thus, review for these services excluded by the contract may also be needed.
Rationale
This evidence review was created in November 1997 and has been updated regularly with searches of the PubMed database. The most recent literature update was performed through Feb. 3, 2022.
Evidence reviews assess whether a medical test is clinically useful. A useful test provides information to make a clinical management decision that improves the net health outcome. That is, the balance of benefits and harms is better when the test is used to manage the condition than when another test or no test is used to manage the condition.
The first step in assessing a medical test is to formulate the clinical context and purpose of the test. The test must be technically reliable, clinically valid, and clinically useful for that purpose. Evidence reviews assess the evidence on whether a test is clinically valid and clinically useful. Technical reliability is outside the scope of these reviews, and credible information on technical reliability is available from other sources.
Computer-Assisted Corneal Topography/Photokeratoscopy
Clinical Context and Test Purpose
The purpose of computer-assisted corneal topography/photokeratoscopy is to provide a diagnostic option that is an alternative to or an improvement on existing therapies, such as manual corneal topography measurements, in patients with disorders of corneal topography.
The question addressed in this evidence review is: Does computer-assisted corneal topography improve health outcomes for patients with disorders of corneal topography, such as keratoconus?
The following PICO was used to select literature to inform this review.
Populations
The relevant population of interest is individuals with disorders of corneal topography.
Interventions
The test being considered is computer-assisted corneal topography/photokeratoscopy.
Comparators
Comparators of interest include manual corneal topography measurements.
Outcomes
The general outcomes of interest are test accuracy, other test performance measures, and functional outcomes.
Identifying clinically validity and usefulness requires short-term follow-up. Evaluating functional outcomes may require longer follow-up.
Study Selection Criteria
Below are selection criteria for studies to assess whether a test is clinically valid.
- The study population represents the population of interest. Eligibility and selection are described.
- The test is compared with a credible reference standard.
- If the test is intended to replace or be an adjunct to an existing test; it should also be compared with that test.
- Studies should report sensitivity, specificity, and predictive values. Studies that completely report true- and false-positive results are ideal. Studies reporting other measures (e.g., receiver operating characteristic, area under receiver operating characteristic, c-statistic, likelihood ratios) may be included but are less informative.
- Studies should also report reclassification of diagnostic or risk category.
Review of Evidence
Clinically Valid
A test must detect the presence or absence of a condition, the risk of developing a condition in the future, or treatment response (beneficial or adverse).
Martinez-Abad et al. (2017) sought to determine whether 3 vector parameters: ocular residual astigmatism, topography disparity, and corneal topographic astigmatism (anterior and total) could serve to detect clinical and subclinical keratoconus.3 One hundred sixty-one eyes were studied in this retrospective comparative study; 61 eyes (38 patients) with keratoconus; 19 eyes (16 patients) with subclinical keratoconus; and a control group of 100 healthy eyes. All study participants underwent a thorough eye exam; further, software was used (iASSORT) to calculate ocular residual astigmatism, topography disparity, and corneal topographic astigmatism. Using a receiver operating characteristic curve analysis, the diagnostic capabilities of the 3 parameters were measured; to further assess diagnostic ability, a cutoff was determined that correlated to the highest sensitivity and specificity of the curve. Results showed that ocular residual astigmatism and topography disparity had good diagnostic capability to detect keratoconus (ocular residual astigmatism: cutoff, 1.255 diopters; sensitivity: 82%; specificity: 92%; topography disparity: cutoff, 1.035 diopters; sensitivity, 78.5%; specificity, 86%). Corneal topographic astigmatism did not show potential as a diagnostic tool.
Section Summary: Clinically Valid
One study has been identified evaluating computer-assisted corneal topography as a clinically valid solution for diagnosing disorders of corneal topography. In it, authors concluded that topography disparity and ocular residual astigmatism, 2 vector parameters that could serve to detect clinical and subclinical keratoconus, were beneficial tools for detecting the disorder.
Clinically Useful
A test is clinically useful if use of the results informs management decisions that improve the net health outcome of care. The net health outcome can be improved if patients receive correct therapy, more effective therapy, or avoid unnecessary therapy or testing.
Direct Evidence
Direct evidence of clinical utility is provided by studies that have compared health outcomes for patients managed with and without the test. Because these are intervention studies, the preferred evidence would be from RCTs.
Contact Lens Fitting
In a study of computer-assisted corneal topography, Bhatoa et al. (2010) assessed the design of gas-permeable contact lens in 30 patients with keratoconus who were recruited in 2005 and 2006.4 The report indicated that the subjects were consecutive, although patients whose topographic plots could not be used were excluded (number not described). The fit of the new lens was compared with the fit of the patient’s habitual lens (randomized order on the same day). Clinical evaluation showed a good fit (no or minor modification needed) for more than 90% of the computer-designed lens. However, the progression of keratoconus causes a bias favoring the most recently fitted lens, confounding comparison between the new computer-designed lens and the patient’s habitual lens. Trial design and reporting limitations limit conclusions that can be drawn from this study.
Weber et al. (2016) reported on a prospective, observational study evaluating the association between computer-assisted corneal topography measurements (Pentacam) and scleral contact lens fit.5 The study included 47 patients (63 eyes) with a variety of indications for scleral contact lenses, most commonly (n = 24 eyes) keratoconus. Pentacam measurements correlated with a subset of the scleral contact lens parameters (corneal astigmatism, anterior chamber depth, and corneal height; p < .001, not adjusted for multiple comparisons) for the group as a whole.
DeNaeyer et al. (2017) investigated the use of the sMap3D system (Precision Ocular Metrology), which measures the surface of the eye for patients in need of a scleral contact lens fitting.6 The sMap3D captures a series of images to produce a single, wide-field topographic “stitched” image of all captured images. To create these images, the patient is asked to provide several “gazes” (gaze up, gaze down, gaze straight). Twenty-five eyes (from 23 patients) were examined using the sMap3D. The “stitched” image produced by the sMap3D was then compared with the single captured straight-gaze image. At a diameter of 10 mm from the corneal center, both straight-gaze image and the sMap3D stitched image displayed 100% coverage of the eye. However, at 14 mm, the straight-gaze image only mapped 68% of the eye; at 15 mm, 53%; at 16 mm, 39%, and at 20 mm, 6%. For the stitched image produced by sMap3D: at 14 mm, 98% coverage; at 15 mm, 96% coverage; at 16 mm, 93% coverage; and at 20 mm, 32% coverage. While there was a significant drop off in coverage between 16 mm and 20 mm for the sMap3D image, the stitched image was considerably more accurate than the straight-gaze image. Tables 2 and 3 provide a summary of the above study characteristics and results.
Bandlitz et al. (2017) studied the profile of the limbal sclera in 8 meridians by using spectral domain optical coherence tomography and a confocal scanning laser ophthalmoscope.7 The objective of this study was to evaluate the relationship between central corneal radii, corneal eccentricity, and scleral radii to improve soft and scleral contact lenses. The limbal scleral radii of 30 subjects were measured. Eight meridians, each 45° apart, were scanned, and it was determined that corneal eccentricity and scleral radii did not correlate in any of the meridians. The authors concluded that the independence between meridians might prove useful in fitting soft and scleral contact lenses.
Table 2. Summary of Key Nonrandomized Study Characteristics
Study | Study Type | Country | Dates | Participants | Treatment 1 | Treatment 2 | Follow- Up |
---|---|---|---|---|---|---|---|
Bhatoa et al. (2010)4 | Randomized, prospective | U.K. | 2005 – 2006 | Patients with keratoconus (n = 30) | Gas-permeable contact lenses made using Fitscan RGP fitting software | Patients habitual RGP contact lenses | NR |
Weber et al. (2016)5 | Prospective, observational | Brazil | 2013 | Patients with a variety of indications for scleral contact lenses (n = 47 patients, 63 eyes) | Pentacam derived topography variables for SCL fit | NA | NR |
DeNaeyer et al. (2017)6 | Retrospective | U.S. | 2016 | Patients presenting for scleral lens fitting (n = 23 patients, 25 eyes) | sMap3D stitched imaging | Straight-gaze imaging | NR |
NA: not applicable; NR: not reported; RGP: rigid gas permeable; SCL: scleral contact lens; U.K.: United Kingdom; U.S.: United States.
Table 3. Summary of Key Nonrandomized Study Results
Study | Agreement Levels between Techniques | Correlations between SCL Parameters and ACD and Hm | Eye Coverage at 10, 14, 16, and 20 mm |
---|---|---|---|
Bhatoa et al. (2010)4 | 74% to 100% | ||
Weber et al. (2016)5 | p < .001, each | ||
DeNaeyer et al. (2017)6 | |||
Straight-gaze | 100%, 68%, 39%, 6% | ||
Stitched | 100%, 98%, 93%, 32% |
ACD: anterior chamber depth; Hm: Pentacam-measured corneal height; SCL: scleral contact lens.
Corneal Astigmatism Measurements for Toric Intraocular Lens Implantation
Lee et al. (2012) reported on a prospective comparative study of 6 methods for measuring corneal astigmatism to guide toric intraocular lens implantation.8 Astigmatism was evaluated in 257 eyes (141 patients) using manual keratometry, auto keratometry, partial coherence interferometry (IOLMaster), ray-tracing aberrometry (iTrace), scanning-slit topography (Orbscan), and Scheimpflug imaging (Pentacam). Each instrument's measurements were masked to the results for the other instruments. The study found no significant difference between instruments, indicating no advantage to computerized corneal topography over manual keratometry.
De Sanctis et al. (2017) reported on corneal astigmatism in patients seeking toric intraocular lens implantation.9 The authors compared 2 methods for measuring corneal astigmatism: (1) corneal astigmatism total corneal refractive power, which uses a ray-tracing method that sends light through the cornea; and (2) corneal astigmatism simulated keratometry, which is a surface-based exterior measurement that measures the steep radius of the anterior cornea. Both methods relied on the camera system (Pentacam HR) to calculate vector differences. Of 200 patients, 77 individuals (60 eyes) remained for intraocular lens implantation. For a patient to qualify for toric intraocular lens implantation, corneal astigmatism had to be greater than 1 diopter. Using corneal astigmatism total corneal refractive power, 17 eyes were found with greater than 1 diopter; using corneal astigmatism simulated keratometry, 13 eyes were found with greater than 1 diopter. However, of the 77 intraocular lens implantation candidates, the corneal astigmatism simulated keratometry method assessed 17 patients to have corneal astigmatism less than or equal to 1 diopter. Moreover, the corneal astigmatism simulated keratometry method found 13 of 123 patients who were not candidates for implantation to have astigmatism greater than 1 diopter. This difference suggested potential issues with patient selection criteria.
Chain of Evidence
Indirect evidence on clinical utility rests on clinical validity. As the evidence is insufficient to demonstrate test performance, no inferences can be made about clinical utility.
A chain of evidence would demonstrate that computer-assisted corneal topography can identify individuals with disorders of corneal topography who would not otherwise be identified, that treatments are available for these patients that would not otherwise be given to patients with disorders of corneal topography, and that these treatments improve health outcomes. Therefore, a chain of evidence cannot be created for clinical utility.
Section Summary: Clinically Useful
Direct evidence for the clinical usefulness of computer-assisted corneal topography in diagnosing those with disorders of corneal topography is lacking. A chain of evidence for clinical validity provides a chain of evidence on clinical usefulness of this testing.
Summary of Evidence
For individuals who have disorders of corneal topography who receive computer-assisted corneal topography/photokeratoscopy, the evidence includes only a few studies. Relevant outcomes are test accuracy, other test performance measures, and functional outcomes. With the exception of refractive surgery, a procedure not discussed herein, no studies have shown clinical benefit (e.g., a change in treatment decisions) based on a quantitative evaluation of corneal topography. In addition, a large prospective series found no advantage with use of different computer-assisted corneal topography methods over manual corneal keratometry. Computer-assisted corneal topography lacks evidence from appropriately constructed clinical trials that could confirm whether it improves outcomes. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.
The purpose of the following information is to provide reference material. Inclusion does not imply endorsement or alignment with the evidence review conclusions.
Practice Guidelines and Position Statements
Guidelines or position statements will be considered for inclusion in Supplemental Information if they were issued by, or jointly by, a U.S. professional society, an international society with U.S. representation, or National Institute for Health and Care Excellence (NICE). Priority will be given to guidelines that are informed by a systematic review, include strength of evidence ratings, and include a description of management of conflict of interest.
American Academy of Ophthalmology
A 1999 American Academy of Ophthalmology (AAO) assessment indicated that computer-assisted corneal topography evolved from the need to measure corneal curvature and topography more comprehensively and accurately than keratometry and that corneal topography is used primarily for refractive surgery.10 The corneal astigmatism simulated keratometry AAO assessment indicated several other potential uses: (1) to evaluate and manage patients following penetrating keratoplasty, (2) to plan astigmatic surgery, (3) to evaluate patients with unexplained visual loss and document visual complications, and (4) to fit contact lenses. However, the corneal astigmatism simulated keratometry AAO assessment noted the lack of data supporting the use of objective measurements (as opposed to subjective determinants, like subjective refraction) of astigmatism.
U.S. Preventive Services Task Force Recommendations
Not applicable
Ongoing and Unpublished Clinical Trials
A search of ClinicalTrials.gov in February 2022 did not identify any ongoing or unpublished trials that would likely influence this review.
References
- Morrow GL, Stein RM. Evaluation of corneal topography: past, present and future trends. Can J Ophthalmol. Aug 1992; 27(5): 213-25. PMID 1393805
- Wilson SE, Klyce SD. Advances in the analysis of corneal topography. Surv Ophthalmol. Jan-Feb 1991; 35(4): 269-77. PMID 2011820
- Martinez-Abad A, Pinero DP, Ruiz-Fortes P, et al. Evaluation of the diagnostic ability of vector parameters characterizing the corneal astigmatism and regularity in clinical and subclinical keratoconus. Cont Lens Anterior Eye. Apr 2017; 40(2): 88-96. PMID 27931882
- Bhatoa NS, Hau S, Ehrlich DP. A comparison of a topography-based rigid gas permeable contact lens design with a conventionally fitted lens in patients with keratoconus. Cont Lens Anterior Eye. Jun 2010; 33(3): 128-35. PMID 20053579
- Weber SL, Ambrosio R, Lipener C, et al. The use of ocular anatomical measurements using a rotating Scheimpflug camera to assist in the Esclera(R) scleral contact lens fitting process. Cont Lens Anterior Eye. Apr 2016; 39(2): 148-53. PMID 26474924
- DeNaeyer G, Sanders DR, Farajian TS. Surface coverage with single vs. multiple gaze surface topography to fit scleral lenses. Cont Lens Anterior Eye. Jun 2017; 40(3): 162-169. PMID 28336224
- Bandlitz S, Baumer J, Conrad U, et al. Scleral topography analysed by optical coherence tomography. Cont Lens Anterior Eye. Aug 2017; 40(4): 242-247. PMID 28495356
- Lee H, Chung JL, Kim EK, et al. Univariate and bivariate polar value analysis of corneal astigmatism measurements obtained with 6 instruments. J Cataract Refract Surg. Sep 2012; 38(9): 1608-15. PMID 22795977
- de Sanctis U, Donna P, Penna RR, et al. Corneal Astigmatism Measurement by Ray Tracing Versus Anterior Surface-Based Keratometry in Candidates for Toric Intraocular Lens Implantation. Am J Ophthalmol. May 2017; 177: 1-8. PMID 28185842
- Corneal topography. American Academy of Ophthalmology. Ophthalmology. Aug 1999; 106(8): 1628-38. PMID 10442914
Coding Section
Codes | Number | Description |
CPT | 92025 | Computerized corneal topography, unilateral or bilateral, with interpretation and report |
92002-92014 | General ophthalmological services | |
HCPCS | No code | |
ICD-10-CM | Not medically necessary for all diagnoses | |
H16.001-H16.9 | Keratitis code range | |
H17.00-H17.9 | Corneal scars and opacities code range | |
H18.001-H18.9 | Other disorders of cornea code range | |
ICD-10-PCS | ICD-10-PCS codes are only used for inpatient services. There is no specific ICD-10-PCS code for this examination. | |
08J0XZZ, 08J1XZZ | Eye examination, code by body part (right eye or left eye) | |
Type of Service | Ophthalmology | |
Place of Service | Physician's Office |
Procedure and diagnosis codes on Medical Policy documents are included only as a general reference tool for each policy. They may not be all-inclusive.
This medical policy was developed through consideration of peer-reviewed medical literature generally recognized by the relevant medical community, U.S. FDA approval status, nationally accepted standards of medical practice and accepted standards of medical practice in this community, Blue Cross Blue Shield Association technology assessment program (TEC) and other nonaffiliated technology evaluation centers, reference to federal regulations, other plan medical policies, and accredited national guidelines.
"Current Procedural Terminology © American Medical Association. All Rights Reserved"
History From 2013 Forward
11/16/2022 | Annual review, no change to policy intent. Updating rationale and references.) |
11/1/2021 |
Annual reivew, no change to policy intent. Updating rationale. |
11/02/2020 |
Annual review, no change to policy intent. Updating regulatory status, guidelines, coding, rationale and references. |
11/01/2019 |
Annual review no change to policy intent. Updating regulatory status and rationale. |
11/28/2018 |
Annual review, no change to policy intent. Updating rationale and references. |
11/01/2017 |
Annual review, no change to policy intent. Updating background, description, rationale and references. |
11/01/2016 |
Annual review, no change to policy intent. Updating background, description, rationale and references. |
11/04/2015 |
Annual review, no change to policy intent. Updating background and description. |
11/12/2014 |
Annual review, no change to policy intent. Added policy guidelines and coding. |
11/06/2013 |
Added Regulatory Status and Benefit Applications. Updated Rationale and References. |