Implantable Bone-Conduction and Bone-Anchored Hearing Aids - CAM 70103

Description:
Sensorineural, conductive, and mixed hearing loss may be treated with various devices, including conventional air-conduction (AC) or bone-conduction external hearing aids. AC hearing aids may not be suitable for patients with chronic middle ear and ear canal infections, atresia of the external canal, or an ear canal that cannot accommodate an ear mold. Bone-conduction hearing aids function by transmitting sound waves through the bone to the ossicles of the middle ear and may be useful for individuals with conductive hearing loss, or (if used with contralateral routing of signal), for unilateral sensorineural hearing loss. Implantable, bone-anchored hearing aids (BAHAs) that use a percutaneous or transcutaneous connection to a sound processor have been investigated as alternatives to conventional bone-conduction hearing aids for patients with conductive or mixed hearing loss or for patients with unilateral single-sided sensorineural hearing loss.

For individuals who have conductive or mixed hearing loss who receive an implantable bone-anchored hearing device with a percutaneous abutment or a partially implantable bone-anchored hearing device with transcutaneous coupling to the sound processor, the evidence includes observational studies that report pre-post differences in hearing parameters after treatment with BAHAs. Relevant outcomes are functional outcomes, quality of life, and treatment-related morbidity. No prospective trials were identified. Observational studies reporting on within-subjects changes in hearing have generally reported hearing improvements with the devices. Given the objectively measured outcomes and the largely invariable natural history of hearing loss in individuals who would be eligible for an implantable bone-conduction device, the demonstrated improvements in hearing after device placement can be attributed to the device. Studies of partially implantable bone-anchored devices have similarly demonstrated within-subjects improvements in hearing. The single-arm studies have shown improvements in hearing in the device-aided state. No direct comparisons other than within-individual comparisons with external hearing aids were identified, but, for individuals unable to wear an external hearing aid, there may be few alternative treatments. The evidence is sufficient to determine qualitatively that the technology results in a meaningful improvement in the net health outcome.

For individuals who have unilateral sensorineural hearing loss who receive a fully or partially implantable bone-anchored hearing device with contralateral routing of signal, the evidence includes 1 randomized controlled trial (RCT), multiple prospective and retrospective case series, and a systematic review. Relevant outcomes are functional outcomes, quality of life, and treatment-related morbidity. Single-arm case series, with sample sizes ranging from 9 to 145 patients, generally have reported improvements in patient-reported speech quality, speech perception in noise, and satisfaction with bone conduction devices with contralateral routing of signal. However, a well-conducted systematic review of studies comparing bone-anchored devices to hearing aids with contralateral routing of signal found no evidence of improvement in speech recognition or hearing localization. The single RCT included in the systematic review was a pilot study enrolling only 10 patients and, therefore, does not provide definitive evidence. The evidence is insufficient to determine the effects of the technology on health outcomes.

For patients with single-sided sensorineural deafness, a binaural hearing benefit may be provided by way of contralateral routing of signals to the hearing ear. There is evidence that bilateral hearing assistance devices improve hearing to a greater degree than unilateral devices. BAHAs may be considered an alternative to external devices in patients who are not candidates for external devices. By extension, use of an implantable bone-conduction device with contralateral routing of signal may be considered medically necessary in patients with unilateral sensorineural deafness.

Background
Hearing Loss
Hearing loss is described as conductive, sensorineural, or mixed, and can be unilateral or bilateral. Normal hearing detects sound at or below 20 decibels (dB). The American Speech-Language-Hearing Association has defined degree of hearing loss based on pure-tone average detection thresholds as mild (20 – 40 dB), moderate (40 – 60 dB), severe (60 – 80 dB), and profound (≥ 80 dB). Pure-tone average is calculated by averaging hearing sensitivities (i.e., the minimum volume that a patient hears) at multiple frequencies (perceived as pitch), typically within the range of 0.25 to 8 kHz.

Sound amplification using an air-conduction (AC) hearing aid can provide benefit to patients with sensorineural or mixed hearing loss. Contralateral routing of signal (CROS) is a system in which a microphone on the affected side transmits a signal to an AC hearing aid on the normal or less affected side.

Treatment
External bone-conduction hearing devices function by transmitting sound waves through the bone to the ossicles of the middle ear. The external devices must be applied close to the temporal bone, with either a steel spring over the top of the head or a spring-loaded arm on a pair of spectacles. These devices may be associated with pressure headaches or soreness.

A bone-anchored implant system combines a vibrational transducer coupled directly to the skull via a percutaneous abutment that permanently protrudes through the skin from a small titanium implant anchored in the temporal bone. The system is based on osseointegration through which living tissue integrates with titanium in the implant over 3 to 6 months, conducting amplified and processed sound via the skull bone directly to the cochlea. The lack of intervening skin permits the transmission of vibrations at a lower energy level than required for external bone-conduction hearing aids. Implantable bone-conduction hearing systems are primarily indicated for people with conductive or mixed sensorineural or conductive hearing loss. They may also be used with CROS as an alternative to an AC hearing aid for individuals with unilateral sensorineural hearing loss.

Partially implantable magnetic bone-conduction hearing systems also referred to as transcutaneous bone-anchored systems, are an alternative to bone-conduction hearing systems that connect to bone percutaneously via an abutment. With this technique, acoustic transmission occurs transcutaneously via magnetic coupling of the external sound processor and the internally implanted device components. The bone-conduction hearing processor contains magnets that adhere externally to magnets implanted in shallow bone beds with the bone-conduction hearing implant. Because the processor adheres magnetically to the implant, there is no need for a percutaneous abutment to physically connect the external and internal components. To facilitate greater transmission of acoustics between magnets, skin thickness may be reduced to 4 to 5 mm over the implant when it is surgically placed.

Regulatory Status
Several implantable bone-conduction hearing systems have been approved by the U.S. Food and Drug Administration for marketing through the 510(k) process (Table 1).

Product codes: MAH, LXB

Table 1. Implantable Bone-Conduction Hearing Systems Approved by the FDA

Device

Manufacturer

Date Cleared

510(k) No.

Baha® Auditory Osseointegrated ImplantSystem

Cochlear Americas

   

BA310 Abutment, BIA310 Implant/Abutment

 

Dec 2018

K182116

Baha 5 Power Sound Processor

 

May 2016

K161123

Baha 5 Super Power Sound Processor

 

Mar 2016

K153245

Baha® 5 Sound Processor

 

Mar 2015

K142907

Baha® Attract System

 

Nov 2013

K131240

Baha® Cordelle II

 

Jul 2015

Apr 2008

K150751

K080363

Baha Divino®

 

Aug 2004

K042017

Baha Intenso® (digital signal processing)

 

Aug 2008

K081606

Baha® 4 (upgraded from the BP100)

 

Sep 2013

K132278

OBC Bone-Anchored Hearing Aid System

Oticon Medical

Nov 2011

K112053

Ponto Bone-Anchored Hearing System

Oticon Medical

Sep 2012

K121228

Ponto 4

 

May 2019

 

Ponto 3, Ponto 3 Power and Ponto 3 SuperPower

 

Sep 2016

K161671

The FDA cleared these systems for use in children ages 5 years and older and adults for the following indications: 

  • Patients who have conductive or mixed hearing loss and can still benefit from sound amplification;
  • Patients with bilaterally symmetric conductive or mixed hearing loss may be implanted bilaterally;
  • Patients with sensorineural deafness in 1 ear and normal hearing in the other (i.e., single-sided deafness);
  • Patients who are candidates for an AC CROS hearing aid but who cannot or will not wear an AC CROS device.

Baha sound processors can be used with the Baha® Softband™. With this application, there is no implantation surgery. The sound processor is attached to the head using a hard or soft headband. The amplified sound is transmitted transcutaneously to the cochlea via the bones of the skull. In 2002, the Baha® Softband™ was cleared for marketing by FDA for use in children younger than 5 years. Because this application has no implanted components, it is not addressed in this evidence review.

The FDA also cleared 2 partially implantable magnetic bone-conduction devices for marketing through the 510(k) process (Table 2)

Table 2. Partially Implantable Magnetic Bone-Conduction Devices Approved by the FDA 

Device

Manufacturer

Date Cleared

510(k) No.

Bonebridge

MED-EL

Mar 2019

K183373

Otomag® Bone-Conduction Hearing System

Medtronic (Formerly Sophono)

Nov 2013

K132189

Cochlear Baha® 4 Sound Processor

Cochlear Americas

Oct 2012

K121317

The SoundBite™ Hearing System (Sonitus Medical, San Mateo, CA) is an intraoral bone-conducting hearing prosthesis that consists of a behind-the-ear microphone and an in-the-mouth hearing device. In 2011, it was cleared for marketing by FDA through the 510(k) process for indications similar to the Baha. However, the manufacturer, Sonitus Medical, closed in 2015.

FDA product code (for bone-anchoring hearing aid): LXB. FDA product code (for implanted bone-conduction hearing aid): MAH.

Related Policies
70105 Cochlear Implant
70184 Semi-Implantable and Fully Implantable Middle Ear Hearing Aids

Policy 
Unilateral or bilateral fully or partially implantable bone-conduction (bone-anchored) hearing aid(s) may be considered MEDICALLY NECESSARY as an alternative to an air-conduction hearing aid in patients 5 years of age and older with conductive or mixed hearing loss who also meet at least one of the following medical criteria:

  • Congenital or surgically induced malformations (e.g., atresia) of the external ear canal or middle ear; or
  • Chronic external otitis or otitis media; or
  • Tumors of the external canal and/or tympanic cavity; or
  • Dermatitis of the external canal;

and meet the following audiologic criteria:

  • A pure-tone average bone-conduction threshold measured at 0.5, 1, 2 and 3 kHz of better than or equal to 45 dB (OBC and BP100 devices), 55 dB (Intenso device), or 65 dB (Cordele II device).

For bilateral implantation, patients should meet the above audiologic criteria and have symmetrically conductive or mixed hearing loss as defined by a difference between left- and right-side bone-conduction threshold of less than 10 dB on average measured at 0.5, 1, 2 and 3 kHz (4 kHz for OBC and Ponto Pro), or less than 15 dB at individual frequencies.

An implantable bone-conduction (bone-anchored) hearing aid may be considered MEDICALLY NECESSARY as an alternative to an air-conduction contralateral routing of signal hearing aid in patients 5 years of age and older with single-sided sensorineural deafness and normal hearing in the other ear. The pure-tone average air-conduction threshold of the normal ear should be better than 20 dB measured at 0.5, 1, 2 and 3 kHz.

Other uses of implantable bone-conduction (bone-anchored) hearing aids, including use in patients with bilateral sensorineural hearing loss, are considered INVESTIGATIONAL.

**PLEASE REFER TO SPECIFIC CONTRACT FOR BENEFITS RELATING TO HEARING AIDS**      

Policy Guidelines
In patients being considered for implantable bone-conduction (bone-anchored) hearing aid(s), skull bone quality and thickness should be assessed for adequacy to ensure implant stability. Additionally, patients (or caregivers) must be able to perform proper hygiene to prevent infection and ensure the stability of the implants and percutaneous abutments.

Coding
Please see the Codes table for details.

Benefit Application
BlueCard/National Account Issues
These devices are referred to as Hearing Aid, Bone Conduction in the U.S. Food and Drug Administration (FDA) approval. The FDA review also indicates that this device has substantially equivalent technology as air-conduction hearing aids with digital sound processing. However, beginning December 2005, the Centers for Medicare & Medicaid Services (CMS) began to consider these devices as prosthetics. Thus, Plans need to review contract language in making decisions about classification.

Benefit limitations regarding hearing aids may apply to these devices. Many medical insurance plans do not provide coverage for hearing aids or provide limited coverage.

State or federal mandates (e.g., FEP) may dictate that all devices approved by the FDA may not be considered investigational. However, this policy considers specific applications of an FDA-approved device as investigational. Alternatively, FDA-approved devices may be assessed on the basis of their medical necessity.

Rationale
This evidence review was created in December 1995 and has been updated regularly with searches of the PubMed database. The most recent literature update was performed through Nov. 15, 2021.

Evidence reviews assess the clinical evidence to determine whether the use of a technology improves the net health outcome. Broadly defined, health outcomes are length of life, quality of life, and ability to function, including benefits and harms. Every clinical condition has specific outcomes that are important to patients and to managing the course of that condition. Validated outcome measures are necessary to ascertain whether a condition improves or worsens; and whether the magnitude of that change is clinically significant. The net health outcome is a balance of benefits and harms.

To assess whether the evidence is sufficient to draw conclusions about the net health outcome of a technology, 2 domains are examined: the relevance and the quality and credibility. To be relevant, studies must represent 1 or more intended clinical use of the technology in the intended population and compare an effective and appropriate alternative at a comparable intensity. For some conditions, the alternative will be supportive care or surveillance. The quality and credibility of the evidence depend on study design and conduct, minimizing bias and confounding that can generate incorrect findings. The randomized controlled trial (RCT) is preferred to assess efficacy; however, in some circumstances, nonrandomized studies may be adequate. RCTs are rarely large enough or long enough to capture less common adverse events and long-term effects. Other types of studies can be used for these purposes and to assess generalizability to broader clinical populations and settings of clinical practice.

Bilateral Implantable Bone-Anchored Hearing Aid Devices With a Percutaneous Abutment in Conductive or Mixed Hearing Loss
Clinical Context and Therapy Purpose

The purpose of implantable BAHAs with a percutaneous abutment is to provide a treatment option that is an alternative to or an improvement on existing therapies, such as external hearing aids, in patients with conductive hearing loss (CHL) or mixed hearing loss.

The question addressed in this evidence review is: Do implantable BAHAs with a percutaneous abutment improve the net health outcome for individuals with CHL or mixed hearing loss?

The following PICO was used to select literature to inform this review.

Populations
The relevant population of interest is individuals with CHL or mixed hearing loss.

Interventions
The therapy being considered are implantable BAHAs with a percutaneous abutment.

Comparators
The main comparator of interest is external hearing aids.

Outcomes
The general outcomes of interest are functional outcomes, quality of life, and treatment-related morbidity.

The existing literature evaluating implantable BAHAs with a percutaneous abutment as a treatment for CHL or mixed hearing loss has varying lengths of follow-up. At least 1 year of follow-up is considered necessary to fully observe outcomes.

Study Selection Criteria
Methodologically credible studies were selected using the following principles:

  • To assess efficacy outcomes, comparative controlled prospective trials were sought, with a preference for RCTs.
  • In the absence of such trials, comparative observational studies were sought, with a preference for prospective studies.
  • To assess longer term outcomes and adverse events, single-arm studies that capture longer periods of follow-up and/or larger populations were sought.
  • Studies with duplicative or overlapping populations were excluded.

Review of Evidence
Janssen et al. (2012) conducted a systematic review to assess the outcomes of bilateral versus unilateral BAHA for individuals with bilateral permanent CHL.1 The literature search included studies in all languages published between 1977 and July 2011. Studies were selected if subjects of any age had permanent bilateral CHL and bilateral implanted BAHAs. Outcomes of interest were any subjective or objective audiologic measures, quality of life indicators, or reports of adverse events. Eleven studies met inclusion criteria; all were observational. The studies included a total of 168 patients, 155 of whom had BAHAs and 146 of whom had bilateral devices. In most studies, comparisons between unilateral and bilateral BAHA were intrasubject. Methodologic heterogeneity between studies precluded meta-analysis, therefore, a qualitative review was performed. Results from 3 (of 11) studies were excluded from synthesis because their patients had been included in multiple publications. Adverse events were not an outcome measure of any of the studies. In general, bilateral BAHA provided additional objective and subjective benefit compared with unilateral BAHA. For example, the improvement in tone thresholds associated with bilateral BAHA ranged from 2 to 15 dB, the improvement in speech recognition patterns ranged from 4 to 5.4 dB, and the improvement in the Word Recognition Score ranged from 1% to 8%. These results were based on a limited number of small observational studies consisting of heterogeneous patient groups that varied in age, the severity of hearing loss, etiology of hearing loss, and previous amplification experience.

Examples of individual studies include the following. Bosman et al. (2001) reported on 25 patients who were using bilateral devices.2 The authors found that both speech recognition in noise and directional hearing improved with the second device. Priwin et al. (2004) reported similar findings in 12 patients with bilateral devices.3 A 2005 consensus statement concluded that bilateral devices resulted in binaural hearing with improved directional hearing and improved speech-in-noise scores in those with bilateral CHL and symmetric bone-conduction thresholds.A number of other studies cited in the 2005 consensus statement found benefits similar to those noted by Bosman and by Priwin.2,3 Positive outcomes continue to be reported: Dun et al. (2010)5 identified improvements in the Glasgow Benefit Inventory scores in 23 children, while Ho et al. (2009)6 reported the same benefit in 93 adults.

Section Summary: Bilateral Bone-Anchored Hearing Aid Devices in Conductive or Mixed Hearing Loss
The evidence on bilateral versus unilateral BAHAs for individuals with CHL or mixed hearing loss consists of small observational studies with heterogeneous participants. In general, bilateral BAHAs seem to provide additional objective and subjective benefit compared with unilateral BAHAs.

Partially Implantable Bone-Anchored Hearing Aid Devices With Transcutaneous Coupling
Clinical Context and Therapy Purpose

The purpose of partially implantable BAHAs with transcutaneous coupling to the sound processer is to provide a treatment option that is an alternative to or an improvement on existing therapies, such as external hearing aids, in patients with CHL or mixed hearing loss.

The question addressed in this evidence review is: do partially implantable BAHAs with transcutaneous coupling to the sound processer improve the net health outcome for individuals with CHL or mixed hearing loss?

The following PICO was used to select literature to inform this review.

Populations
The relevant populations of interest is individuals with CHL or mixed hearing loss.

Interventions
The therapy being considered is partially implantable BAHAs with transcutaneous coupling to the sound processor, wherein acoustic transmission occurs transcutaneously via magnetic coupling of an external sound processor to the internally implanted device components.

Comparators
The main comparator of interest is external hearing aids.

Outcomes
The general outcomes of interest are functional outcomes, quality of life, and treatment-related morbidity.

The existing literature evaluating partially implantable BAHAs with transcutaneous coupling to the sound processer as a treatment for CHL or mixed hearing loss has varying lengths of follow-up. At least 1 year of follow-up is considered necessary to fully observe outcomes.

Study Selection Criteria
Methodologically credible studies were selected using the following principles:

  • To assess efficacy outcomes, comparative controlled prospective trials were sought, with a preference for RCTs.
  • In the absence of such trials, comparative observational studies were sought, with a preference for prospective studies.
  • To assess longer term outcomes and adverse events, single-arm studies that capture longer periods of follow-up and/or larger populations were sought.
  • Studies with duplicative or overlapping populations were excluded.

Review of Evidence
Prospective Studies

Two prospective studies (discussed below) evaluating different transcutaneous systems were identified. Both trials were small (27 and 15 individuals), but both demonstrated improvements in hearing outcomes.

Briggs et al. (2015) reported on a prospective interventional evaluation of the percutaneous, partially implantable Baha Attract System among 27 adults with CHL or mild mixed hearing loss in the ear to be implanted.7 The choice of sound processor was based on patient preference and hearing tests with various sound processors in conjunction with the Baha Softband before device implantation. All 27 patients enrolled received an implant. Sound processor fitting occurred 4 weeks postimplantation in all but 1 patient. At 9-month follow-up, pure-tone audiometry (PTA; means of 500, 1000, 2000 and 4000 Hz) was significantly improved with the implant and sound processor compared with unaided hearing (18.4-dB hearing loss; p < 0.001). Patients generally showed improvements in speech recognition in noise, although comparing results across test sites was difficult due to different languages and methodologies used for testing speech recognition at each site. Compared with the preoperative unaided state, scores on the Abbreviated Profile of Hearing Aid Benefit overall score (p = 0.038) and reverberation (p = 0.016) and background noise (p=0.035) subscales were significantly improved with the test device.

Denoyelle et al. (2015) reported on a prospective trial of the Sophono device in children ages 5 to 18 years with uni- or bilateral congenital aural atresia with complete absence of the external auditory canal with pure CHL.8 The study included a within-subject comparison of hearing results with the Sophono devices to those obtained with the Baha Softband preoperatively. All 15 patients enrolled were implanted (median age, 97 months). At 6-month follow-up, mean aided AC PTA was 33.49 dB (mean gain, 35.53 dB), with a mean aided sound reception threshold of 38.2 dB (mean gain, 33.47 dB). The difference in AC PTA between the Baha Softband and the Sophono device was 0.6 dB (confidence interval upper limit, 4.42 dB), which met the trial’s prespecified noninferiority margin. Adverse events were generally mild, including skin erythema in 2 patients, which improved by using a weaker magnet, and brief episodes of pain or tingling in 3 patients.

Nonrandomized Comparative Studies
A limited amount of data is available comparing transcutaneous with percutaneous bone-anchored conduction devices. Hol et al. (2013) compared percutaneous BAHA implants with partially implantable magnetic transcutaneous bone-conduction hearing implants using the Otomag Sophono device in 12 pediatric patients (age range, 5 to 12 years) who had congenital unilateral CHL.9 Sound-field thresholds, speech recognition threshold, and speech comprehension at 65 dB were somewhat better in patients with the BAHA implant (n = 6) than in those with the partially implantable hearing device (n = 6). Using a skull simulator, output was 10 to 15 dB lower with the partially implantable device than with the BAHA device. After following the same 12 patients for more than 3 years, Nelissen et al. (2016) reported on soft tissue tolerability, hearing results, and sound localization abilities.10 Two patients in each group had stopped using their hearing devices. Soft tissue tolerability with the Sophono device was favorable compared with BAHA. Both groups showed improvements in sound localization compared with the unaided situation. Aided thresholds with the Sophono were not as good as expected, with a mean pure-tone average of about 30 dB hearing loss; ideally aided thresholds should be 10 to 20 dB hearing loss.

Iseri et al. (2015) described a retrospective, single-center study from Turkey comparing 21 patients treated using a transcutaneous, fully implantable BAHA with 16 patients treated using a percutaneous device (the Baha Attract).11 Groups were generally similar at baseline, with most individuals undergoing BAHA placement for chronic otitis media. Operating time was longer in patients treated with the transcutaneous partially implantable devices (46 minutes vs. 26 minutes, p < 0.05). Three patients treated with percutaneous devices had Holgers grade 2 skin reactions, and 2 stopped using their devices for reasons unrelated to skin reactions. Mean thresholds for frequencies 0.5 to 4.0 kHz were 64.4 dB without the BAHA and 31.6 dB with the BAHA in the percutaneous device group, and 58.3 dB without the BAHA and 27.2 dB with the BAHA in the transcutaneous device group. Frequency-specific threshold hearing gains did not differ significantly between groups. Mean hearing gain measured by speech reception threshold was statistically significantly smaller in the percutaneous group (24 dB vs. 36.7 dB, p = 0.02).

Gerdes et al. (2016) published a retrospective single-center study comparing 10 patients who had CHL who received the transcutaneous Bonebridge device with an audiologically matched control group of 10 patients who received the percutaneous BAHA BP100.12 There were similar significant improvements in aided thresholds, word recognition scores, and speech reception thresholds in noise for both devices. There were also no differences in subjective ratings for the Abbreviated Profile of Hearing Aid Benefit. Mean functional gain was slightly higher (27.5 dB) for transcutaneous than for percutaneous (26.3 dB), but not significantly different.

Observational Studies
Dimitriadis et al. (2016) reported on a systematic review of observational studies of the BAHA Attract device including 10 studies (N = 89 patients; range, 1 to 27 patients).13, Seventeen (19%) of the patients were children, of whom 5 had unilateral sensorineural hearing loss and 4 had CHL. Of the 27 (45%) adults, 22 had unilateral sensorineural hearing loss, and 11 (18%) had bilateral mixed hearing loss. Audiologic and functional outcome measures and the timing of testing varied greatly in the studies. Summary measures were not reported. In general, audiologic and functional outcomes measured pre- and post-implantation showed improvement, although statistical comparisons were lacking in some studies.

Reddy-Kolanu et al. (2016) reported on complications with the BAHA Attract (n = 34) from a case series that included all patients implanted in a single center between 2013 and 2015.14 Patients ranged in age from 8 to 64 years, and follow-up ranged from 3 to 20 months. Twenty-three patients had no significant postoperative problems. Five patients required an alteration in magnet strength primarily due to implant site tenderness. One patient reported distressing tinnitus; another had the implant changed to an abutment system due to infection, and a third had the magnet removed following trauma to the implant site. One patient has ongoing psoriasis problems. Two patients were converted to a newer, lighter sound processor.

In an early (2011) study, Seigert reported on the use of a transcutaneous, partially implantable bone-conduction hearing system (Otomag).15 Among 12 patients who received the system, there were average hearing gains of 31.2 dB in free-field PTA. The free-field suprathreshold speech perception at 65 dB increased from 12.9% preimplantation to 72.1% postimplantation.

Powell et al. (2015) reported on outcomes from a retrospective study that included 6 patients treated with the Otomag Sophono device and 6 treated with the BAHA Attract device.16 Ten subjects were identified as the primary author’s patients and the remaining were identified through an Australian national hearing database. In the BAHA Attract group, mean AC thresholds across 4 frequencies (0.5, 1, 2 and 4 kHz) improved from 60.8 dB in the unaided state to 30.6 dB in the aided state. In the Sophono group, the mean 4-frequency AC thresholds improved from 57.8 dB in the unaided state to 29.8 dB in the aided state. Speech discrimination in noise scores did not differ significantly between devices.

O’Niel et al. (2014) reported on outcomes for 10 pediatric patients with CHL treated with the Otomag Sophono device at a single center.17 Fourteen ears were implanted with no surgical complications. The skin complication rate was 35.7%, including skin breakdown (n = 2) and pain and erythema (n = 5); negative outcomes resulted in 5 (36%) of 14 ears having sufficient difficulties to discontinue device use for a period. Mean aided PTA was a 20.2-dB hearing level, with a mean functional gain of a 39.9-dB hearing level. Patients without skin complications consistently used their devices (average daily use, 8 to 10 hours).

Centric et al. (2014) also reported on outcomes for 5 pediatric patients treated with the Otomag Sophono device at a single center.18 Etiologies of hearing loss were heterogeneous and included bilateral moderate or severe CHL and unilateral sensorineural hearing loss. The average improvement in PTA was a 32-dB hearing level, and the average improvement in speech response threshold was a 28-dB hearing level. All patients responded in the normal-to-mild hearing loss range in the implanted ear after device activation. In a follow-up study from the same institution, Baker et al. (2015) reported pooled outcomes for the first 11 patients treated with the Otomag Sophono and the first 6 patients treated with the Baha Attract.19 Pre- and post-implant audiometric data were available for 11 ears in the Sophono group and 5 in the Baha Attract group. Average improvement over all frequencies ranged from a 24- to 43-dB hearing level in the Sophono group and from a 32- to 45-dB hearing level in the Baha Attract group. The average improvement in PTA was a 38-dB hearing level in the Sophono group and a 41-dB hearing level in the Baha Attract group.

Other single-center observational series have described clinical experience with transcutaneous partially implantable BAHA devices. Marsella et al. (2014) reported on outcomes for 6 pediatric patients treated with the Otomag Sophono device for CHL or mixed hearing loss.20 Median improvement in PTA was 33-dB hearing loss, and median free-field PTA (0.5 to 3 kHz) with the device was 32.5-dB hearing loss. Magliulo et al (2015) reported on outcomes for 10 patients treated with the Otomag Sophono device after subtotal petrosectomy for recurrent chronic middle ear disease, a procedure associated with a CHL of 50 to 60 dB.21, Postsurgery with the Sophono device, there was an average acoustic improvement in AC of 29.7 dB, which was significantly better than the improvement seen with traditional AC hearing aids (18.2 dB).

In addition to studies of partially implantable bone-conduction devices currently approved by the U.S. Food and Drug Administration (FDA), a number of case series identified evaluated the Bonebridge implant, which was recently cleared for marketing in the United States in March 2019. Case series with at least 5 patients are summarized in Table 3.

Table 3. Case Series Evaluating the Bonebridge Implant

Study N Patient Population Main Hearing Results Safety Outcomes
Bravo-Torres et al. (2018)22 15
  • Pediatric patients with bilateral CHL (microtia associated with external auditory canal atresia)
  • Aided sound-field threshold improvement: 25.2 dB
Minor feedback (4), broken processors (4), mild skin redness (2) with 1 month follow-up
Schmerber et al. (2017)23 25
  • SSD (n = 12)
  • Bilateral CHL (n = 7)
  • Bilateral mixed HL (n = 6)
  • SSD, in 5/7 patients speech reception threshold in noise lower with Bonebridge activated
  • CHL and mixed, average functional gain: 26 dB HL; mean % of speech recognition in quiet improved from 74% unaided to 95% aided
No complications, device failures, revision surgery, or skin injury reported with 1 y follow-up
Rahne et al. (2015)24 11
  • SSD (n = 6; 1 sensorineural, 3 mixed, 2 conductive)
  • Bilateral CHL (n = 2)
  • Bilateral mixed HL or mixed/sensorineural (n = 3)
  • Aided sound-field threshold improvement: 33.4 dB
  • WRS improved from mean of 10% unaided to 87.5% aided
1 case of chronic fibrosing mastoiditis requiring mastoidectomy and antrotomy; no other complications
Laske et al. (2015)25 9
  • Adults with SSD and normal contralateral hearing
  • Speech discrimination signal-to-noise improvement for aided vs unaided condition, sound presented to aided ear: 1.7 dB
  • Positive improvements on quality-of-life questions
Not reported
Riss et al. (2014)26 24
  • Combined HL (n = 9)
  • EAC atresia (n = 12)
  • SSD (n = 3)
  • Average functional gain: 28.8 dB
  • Monosyllabic word scores at 65-dB sound pressure increased from 4.6 to 53.7 percentage points
Not reported
Manrique et al. (2014)27 5
  • Mixed HL (n = 4)
  • SSD (n = 1)
  • PTA improvement: 35.62 dB (p=0.01)
  • Disyllabic word discrimination improvement: 20% (p=0.016)
No perioperative complications reported
Ihler et al. (2014)28 6
  • Mixed HL (n = 4)
  • CHL (n = 2)
  • PTA functional gain (average, 0.5 to 4.0 kHz): 34.5 dB
  • Speech discrimination at 65 dB improvement:
    • In quiet: 63.3 percentage points
    • In noise: 37.5 percentage points
Prolonged wound healing in 1 case
Desmet et al. (2014)29 44
  • All unilaterally deaf adults
  • Statistically significant improvement on APHAB and SHHIA
Not reported
Iseri et al. (2014)30 12
  • CHL (n = 9)
  • “Primarily conductive hearing loss” (n = 3)
  • Speech reception threshold increase: 19 dB
Postoperative hematoma requiring aspiration in 1 case


APHAB: Abbreviated Profile of Hearing Aid Benefit; CHL: conductive hearing loss; EAC: external auditory canal; HL: hearing loss; PTA: pure-tone average; SHHIA: Short Hearing Handicap Inventory for Adults; SSD: single-sided deafness; WRS: Word Recognition Score.

Section Summary: Partially Implantable Magnetic Bone-Anchored Hearing Aid Devices
Studies of transcutaneous, partially implantable BAHAs have typically used a retrospective within-subjects comparison of hearing thresholds with and without the device, although there have been 2 small (27 and 15 participants) prospective studies. There was heterogeneity in the audiologic and functional outcome measures used in the studies and the timing of testing. Studies of partially implantable BAHAs have generally demonstrated within-subjects’ improvements in hearing.

Fully or Partially Implantable Bone-Anchored Hearing Aid Devices with Contralateral Routing of Signal for Unilateral Sensorineural Hearing Loss
Clinical Context and Therapy Purpose

The purpose of fully or partially implantable BAHAs with contralateral routing of signal (CROS) is to provide a treatment option that is an alternative to or an improvement on existing therapies, such as air-conduction hearing aids with CROS, in patients with unilateral sensorineural hearing loss.

The question addressed in this evidence review is: do fully or partially implantable BAHAs with CROS improve the net health outcome for individuals with unilateral sensorineural hearing loss?

The following PICO was used to select literature to inform this review.

Populations
The relevant population of interest is individuals with unilateral sensorineural hearing loss, also called single sided deafness. In this population, 1 ear has minimal to moderate hearing loss while the other ear has significant sensorineural hearing loss. Patients with unilateral sensorineural hearing loss often have difficulty hearing or understanding conversation on their impaired side, particularly in the presence of background noise.

Interventions
The therapy being considered is fully or partially implantable BAHAs with CROS, a system that transmits sound from the affected side to the normal or less affected side.

Comparators
The main comparator of interest is air-conduction hearing aids. Also referred to as acoustic hearing aids, the air-conduction hearing aid is a standard treatment for conductive, mixed, sensorineural, and medically and surgically unresponsive CHL. These aids are rated as Class I by the FDA.

Outcomes
The general outcomes of interest are functional outcomes, quality of life, and treatment-related morbidity.

The existing literature evaluating partially implantable BAHAs with CROS as a treatment for CHL or mixed hearing loss has varying lengths of follow-up. At least 1 year of follow-up is considered necessary to fully observe outcomes.

Study Selection Criteria
Methodologically credible studies were selected using the following principles:

  • To assess efficacy outcomes, comparative controlled prospective trials were sought, with a preference for RCTs
  • In the absence of such trials, comparative observational studies were sought, with a preference for prospective studies.
  • To assess longer term outcomes and adverse events, single-arm studies that capture longer periods of follow-up and/or larger populations were sought.
  • Studies with duplicative or overlapping populations were excluded.

Review of Evidence
Systematic Reviews

Peters et al. (2015) reported results from a systematic review of studies comparing BAHA devices using CROS) systems with hearing aids using CROS for single-sided deafness (SSD).31 Six studies met eligibility criteria, including 1 RCT and 3 prospective and 2 retrospective case series, 5 of which were considered to have moderate-to-high directness of evidence and low-to-moderate risk of bias. The 5 studies (n = 91 patients) with low or moderate risk of bias were noted to have significant heterogeneity in the populations included. For speech perception in noise, there was no consistent improvement with aided hearing over an unaided hearing in all environments. All studies reported equal sound localization and quality of life outcomes for both hearing conditions.

Baguley et al. (2006) reviewed the evidence for contralateral BAHAs in adults with acquired unilateral sensorineural hearing loss.32 None of the 4 controlled trials reviewed showed a significant improvement in auditory localization with the bone-anchored device. However, speech discrimination in noise and subjective measures improved with these devices: the BAHAs resulted in greater improvements than those obtained with the conventional AC CROS systems.

Prospective Studies
Since the publication of the Peters systematic review, 3 prospective, interventional studies have compared patient outcomes using transcutaneous BAHA devices with CROS hearing aids for SSD. den Besten et al. (2018) assessed 54 adults with SSD, each of whom underwent a trial with the Baha Softband before a trial of the percutaneous, partially implantable Baha Attract device.33 No statistically significant difference in audiological outcomes was seen between the 2 devices (p > 0.05). At a 6 month follow-up after implantation, patients reported numbness (20%) and slight pain/discomfort (38%) associated with the device. Leterme et al. (2015) assessed 24 adults with SSD, 18 of whom were evaluated with trials of both hearing aids with CROS and bone-conduction‒assisted hearing using the Baha Softband.34 Most (72%) patients, after completing trials of both devices, preferred the BAHA device to hearing aids with CROS. Glasgow Benefit Inventory and Abbreviated Profile of Hearing Aid Benefit scores did not differ significantly between devices. Sixteen of the 18 subjects elected to undergo implantation of a percutaneous BAHA device. In general, hearing improvement with the Baha Softband trial correlated with hearing improvements following device implantation. Snapp et al. (2017) reported on a prospective single-center study of 27 patients with unilateral severe-profound sensorineural hearing loss who had either a CROS (n = 13) or transcutaneous BAHA (n = 14) device.35 Mean device use was 66 months for the BAHAs and 34 months for CROS devices. Both BAHA and CROS groups had significant improvement in speech-in-noise performance, but neither showed improvement in localization ability. There were no differences between the devices for subjective measures of posttreatment residual disability or satisfaction as measured by the Glasgow Hearing Aid Benefit Profile.

Observational Studies
Zeitler et al. (2012) reported on a retrospective case series of 180 patients with SSD and residual hearing in the implanted ear who underwent unilateral or bilateral BAHA placement at a U.S. university medical center.36 Significant improvement was reported in objective hearing measures (speech-in-noise and monosyllabic word tests) following BAHA implantation. Subjective benefits from BAHAs varied across patients based on results from the Glasgow Hearing Aid Benefit Profile, but patients with residual hearing in the affected ear tended toward improved satisfaction with their device postoperatively.

Additional series from various countries, with sample sizes ranging from 9 to 145 patients, have reported on outcomes after implantation of BAHAs for SSD. In general, these studies have indicated improvements in patient-reported speech quality, speech perception in noise, and patient satisfaction.37,38,39,40,41,42,43,44

Section Summary: Bone-Anchored Hearing Aid Devices for Unilateral Sensorineural Hearing Loss
Single-arm case series with sample sizes ranging from 9 to 180 patients have generally reported some improvements in patient-reported outcomes after implantation of bone-conduction devices, but no improvements in speech recognition or hearing localization. However, in studies with comparators, outcomes for patients with bone-anchored devices were similar to those for patients with hearing aids with CROS.

Bone-Anchored Hearing Aid Devices in Children Younger Than Age 5 Years
The BAHA device has been investigated in children younger than 5 years in Europe. Reports have described experiences with preschool children or children with developmental issues that might interfere with device maintenance and skin integrity. A 2-stage procedure may be used in young children. In the first stage, the fixture is placed into the bone and allowed to fully osseointegrate. After 3 to 6 months, a second procedure is performed to connect the abutment through the skin to the fixture.

The largest series in children under 5 years were identified, described by Amonoo-Kuofi et al. (2015) and including 24 children from a single center’s prospectively maintained database.45 Most patients underwent a 2-stage surgical approach. Most (52%) patients received the implant for isolated microtia or Goldenhar syndrome (16%). Following implantation, 13 (54%) patients had grade 2 or 3 local reactions assessed on the Holgers Classification System (redness, moistness, and/or granulation tissue) and 7 (29%) had grade 4 local reactions on this scale (extensive soft tissue reaction requiring removal of the abutment). Quality of life scores (Glasgow Children’s Benefit Inventory; scoring range, -100 to 100) were obtained in 18 subjects/parents, with a final mean score change of +40 points. Audiologic testing indicated that the average performance of the device fell within the range of normal auditory perception in noisy and quiet environments.

Marsella et al. (2012) reported on a single-center experience in Italy with pediatric BAHAs from the inception of their program in 1995 to December 2009.46 Forty-seven children (21 girls, 26 boys) were implanted; 7 were younger than 5 years. The functional gain was significantly better with BAHAs than with conventional nonimplanted bone-conduction hearing aids, and there was no significant difference regarding functional outcomes between the 7 younger patients and the rest of the cohort. Based on these findings, study authors suggested that implantation of children at an age younger than 5 years can be conducted safely and effectively in such settings. Report conclusions were limited by the small number of very young children in the sample and the limited statistical power to detect a difference between younger and older children.

Davids et al. (2007) provided BAHA devices to children younger than 5 years of age for auditory and speech-language development, and retrospectively compared surgical outcomes for a study group of 20 children younger than 5 years and a control group of 20 older children.47 Children with a cortical bone thickness greater than 4 mm underwent a single-stage procedure. The interstage interval for children having 2-stage procedures was significantly longer in the study group to allow implantation in younger patients without increasing surgical or postoperative morbidity. Two traumatic fractures occurred in the study group versus 4 in the older children. Three younger children required skin site revision. All children were wearing their BAHA devices at the time of writing. McDermott et al. (2008) reported on the role of BAHAs in children with Down syndrome in a retrospective case analysis and postal survey of complication rates and quality of life outcomes for 15 children ages 2 to 15 years.48 All used their BAHA devices at a 14-month follow-up. No fixtures were lost; skin problems were encountered in 3 patients. All 15 patients had improved social and physical functioning, attributed to improved hearing.

Section Summary: Bone-Anchored Hearing Aid Devices in Children Younger Than Age 5 Years
There are few data on the use of BAHA devices in children younger than 5. Three case series with a total of fewer than 60 children younger than 5 years have reported improvements in quality of life after implantation with BAHA devices. One comparative observational study, with 7 children younger than 5, reported significantly better improvement in functional gain with BAHAs than with conventional nonimplanted bone-conduction hearing aids in an analysis including all ages.

Safety and Adverse Events Related to Bone-Anchored Hearing Aid Devices
Review of Evidence
Systematic Reviews

Schwab et al. (2020) completed a systematic review of adverse events associated with bone-conduction and middle-ear implants.49 The 10 most frequently reported adverse events for bone conduction hearing implants included skin reactions (Holgers grade 1 to 3), skin revision surgery due to overgrowth or cellulitis, minor soft tissue/skin overgrowth, skin infection, surgical revision, preimplantation, failure to osseointegrate, and minor skin complications.

Verheij et al. (2016) published a systematic review on complications of surgical tissue preservation techniques with percutaneous BAHA devices including 18 studies with 381 devices.50 The implantation techniques reported in the studies were as follows: punch method, 4 studies (81 implants); linear incision technique without soft tissue reduction, 13 studies (288 implants); and Weber technique, 1 study (12 implants). Indications for surgery were SSD (n = 68), sensorineural hearing loss (n = 4), mixed hearing loss (n = 65), or CHL (n = 66). The Holgers classification was used to grade soft tissue reactions (grade 0, no reaction; grade 2, red and moist tissue; grade 3, granulated tissue; grade 4, removal of skin-penetrating implant necessary due to infection). The incidence of Holgers grade 3 was 2.5% with the punch technique, 5.9% with the linear incision technique, and 0% with the Weber technique. Holgers grade 4 was reported in 1 patient implanted with the linear incision technique.

Kiringoda and Lustig (2013) reported on a meta-analysis of complications related to BAHA implants. Selected were 20 studies that evaluated complications in 2,134 adult and pediatric patients who received a total of 2,310 BAHA implants.51 The quality of available studies was considered poor and lacking in uniformity. Complications related to BAHA implants were mostly minor skin reactions: The incidence of Holgers Classification System grade 2, 3 or 4 skin reactions ranged from 2.4% to 38.1% in all studies. The incidence of failed osseointegration ranged from 0% to 18% in adult and mixed population studies and from 0% to 14.3% in pediatric population studies. The incidence of revision surgery ranged from 1.7% to 34.5% in adult and mixed population studies and from 0.0% to 44.4% in pediatric population studies. Implant loss ranged from 1.6% to 17.4% in adult and mixed population studies and from 0.0% to 25% in pediatric studies.

Observational Studies
Dun et al. (2012) assessed soft tissue reactions and implant stability of 1,132 percutaneous titanium implants for bone-conduction devices in a retrospective survey of 970 patients undergoing implants between 1988 and 2007 at a university medical center in the Netherlands.52 Study investigators also examined device usage and compared different patient age groups (children, adults, elderly patients) over a 5-year follow-up. Implant loss was 8%. In close to 96% of cases, there were no adverse soft tissue reactions. Significantly more soft tissue reactions and implant failures were observed in children than in adults and elderly patients (p < 0.05). Implant survival rates were lower in patients with than without mental retardation (p = 0.001).

Hobson et al. (2010) reviewed complications of 602 patients at a tertiary referral center over 24 years and compared their observed rates with those published in 16 previous studies.53 The overall observed complication rate of 23.9% (144/602) was similar to other published studies (weighted mean complication rate, 24.9%). The most common complications were soft tissue overgrowth, skin infection, and fixture dislodgement. The observed rate of surgical revision of 12.1% (73/602) was also similar to previously published rates (weighted mean, 12.7%). Top reasons for revision surgery were identical to observed complications. Wallberg et al. (2011) reported on the status of 150 implants placed between 1977 and 1986 at a mean follow-up of 9 years.54 Implants were lost in 41 (27%) patients. Reasons for implant loss were: removal (16 patients), osseointegration failure (17 patients), and direct trauma (8 patients). In the 132 patients with implant survival, BAHAs were still being used by 119 (90%) patients at the 9-year follow-up. For children, implant complications were even more frequent, as reported by Kraai et al. (2011) in a follow-up evaluation of 27 implants placed in children ages 16 years or younger between 2002 and 2009.55 In this retrospective report, soft tissue reactions occurred in 24 (89%) patients; implant removal or surgical revision was required in 10 (37%) patients; 24 (89%) patients experienced soft tissue overgrowth and infection; and 7 (26%) patients experienced implant trauma. Chronic infection and overgrowth at the abutment prevented use of the implant in 3 (11%) patients.

Allis et al. (2014) conducted a prospective observational cohort study with a retrospective historical control to evaluate complication rates of skin overgrowth, infection, and the need for revision surgery associated with a BAHA implant with a longer (8.5-mm) abutment.56 Twenty-one subjects were treated with the 8.5-mm abutment implant from 2011 to 2012 and were compared with 23 subjects treated with a 5.5-mm abutment implant from 2010 to 2011. Groups were generally similar at baseline, except that patients with the 8.5-mm abutment implant were older (62 years vs. 48 years, p = 0.012). Patients in the longer abutment group were less likely to experience infection (10% vs. 43%; p = 0.02), skin overgrowth (5% vs. 41%; p = 0.007), and need for revision (10% vs. 45%; p = 0.012), respectively.

Other observational cohort studies, ranging in size from 47 to 974 subjects, have reported safety and adverse event outcomes after BAHA placement.57,58,59,60 Across these studies, implant loss ranged from 4% to 18%.

Different surgical techniques for implanting BAHA devices and specific BAHA designs have yielded better safety outcomes. In a 2016 systematic review of 30 articles on the association between surgical technique and skin complications following BAHA implantation, the dermatome technique (vs. a skin graft or linear technique) was linked to more frequent skin complications.61 Fontaine et al. (2014) compared complication rates for 2 BAHA surgical implantation techniques among 32 patients treated from 2004 to 2011.62 Complications requiring surgical revision occurred in 20% of cases who had a skin flap implantation method (n = 20) and in 38% of cases who had a full-thickness skin graft implantation method (n = 21; p = 0.31). Hultcrantz and Lanis (2014) reported shorter surgical times and fewer cases of numbness and peri-implant infections in 12 patients treated with a non-skin-thinning technique, compared with 24 patients treated with a flap or a dermatome implantation technique.63 In a comparison of 2 types of BAHA devices, 1 with a 4.5-mm diameter implant and a rounded 6-mm abutment (n = 25) and 1 with a 3.75-mm diameter implant and a conically shaped 5.5-mm abutment (n = 52), Nelissen et al. (2014) reported that implant survival was high for both groups over a 3-year follow-up, although the conically shaped abutment had greater stability.64 Singam et al. (2014) reported results of a BAHA implantation technique without soft tissue reduction in conjunction with a longer device abutment in 30 patients.65 Twenty-five patients had no postoperative complications. Five subjects developed postoperative skin reactions, of whom 3 required soft tissue reduction. Roplekar et al. (2016) compared skin-related complications of the traditional skin flap method to the linear incision method performed by a single surgeon in 117 patients with at least 1 year of follow-up.66 Twenty-one (24%) patients experienced skin-related complications in the skin flap group (12 skin overgrowths, 8 wound infections, 1 numbness) and 3 (10%) patients experienced complications in the linear incision group (3 wound infections).

Section Summary: Safety and Adverse Events Related to Bone-Anchored Hearing Aid Devices
The quality of available data for adverse events is generally poor with high heterogeneity. The most frequently reported complications from surgical procedures for BAHA insertion are adverse skin reactions, with an incidence of Holgers grade 2, 3 or 4 reactions ranging from less than 2% to more than 34%, and implant loss ranging from less than 2% to more than 17%. There is some evidence of reductions in complication rates and their severity with newer surgical techniques (e.g., linear incision).

Summary of Evidence
For individuals who have conductive or mixed hearing loss who receive an implantable BAHA with a percutaneous abutment or a partially implantable BAHA with transcutaneous coupling to the sound processor, the evidence includes observational studies that have reported pre-post differences in hearing parameters after treatment with BAHAs. Relevant outcomes are functional outcomes, quality of life, and treatment-related morbidity. No prospective trials were identified. Observational studies reporting on within-subjects changes in hearing have generally reported hearing improvements with the devices. Given the objectively measured outcomes and the largely invariable natural history of hearing loss in individuals who would be eligible for an implantable bone-conduction device, the demonstrated improvements in hearing after device placement can be attributed to the device. Studies of partially implantable BAHAs have similarly demonstrated within-subjects improvements in hearing. The single-arm studies have shown improvements in hearing in the device-aided state. No direct comparisons other than within-individual comparisons with external hearing aids were identified, but, for individuals unable to wear an external hearing aid, there may be few alternative treatments. The evidence is sufficient to determine that the technology results in an improvement in the net health outcome.

For individuals who have unilateral sensorineural hearing loss who receive a fully or partially implantable BAHA with the contralateral routing of signal, the evidence includes an RCT, multiple prospective and retrospective case series, and a systematic review. Relevant outcomes are functional outcomes, quality of life, and treatment-related morbidity. Single-arm case series, with sample sizes ranging from 9 to 180 patients, have generally reported improvements in patient-reported speech quality, speech perception in noise, and satisfaction with bone-conduction devices with contralateral routing of the signal. However, a well-conducted systematic review of studies comparing bone-anchored devices with hearing aids using contralateral routing of signal found no evidence of improvement in speech recognition or hearing localization. The single RCT included in the systematic review was a pilot study enrolling only 10 patients and, therefore, does not provide definitive evidence. Quality RCTs on BAHA for unilateral sensorineural hearing loss are lacking. The evidence is insufficient to determine that the technology results in an improvement in the net health outcome.

The purpose of the following information is to provide reference material. Inclusion does not imply endorsement or alignment with the evidence review conclusions.

Clinical Input From Physician Specialty Societies and Academic Medical Centers
While the various physician specialty societies and academic medical centers may collaborate with and make recommendations during this process, through the provision of appropriate reviewers, input received does not represent an endorsement or position statement by the physician specialty societies or academic medical centers, unless otherwise noted.

2016 Clinical Input
In response to requests, input was received from 2 specialty societies and 3 academic medical centers (1 of which provided 4 responses and 1 of which provided 3 responses) while this policy was under review in 2016. Input focused on the categorization of partially implantable bone-anchored devices relative to fully implantable devices. There was a strong consensus that partially implantable devices are considered an evolution of earlier devices, and that direct trials comparing the 2 are not necessary.

Practice Guidelines and Position Statements
Guidelines or position statements will be considered for inclusion in Supplemental Information if they were issued by, or jointly by, a U.S. professional society, an international society with U.S. representation, or National Institute for Health and Care Excellence (NICE). Priority will be given to guidelines that are informed by a systematic review, include strength of evidence ratings, and include a description of management of conflict of interest.

In 2021 , the American Academy of Otolaryngology Head and Neck Surgery updated its position statement on the use of implantable hearing devices.67 It states that the Academy “considers bone conduction hearing devices (BCHD) as appropriate, and in some cases preferred, for the treatment of conductive and mixed hearing loss. BCHD may also be indicated in select patients with single-sided deafness. BCHD include semi-implantable bone conduction devices utilizing either a percutaneous or transcutaneous attachment, as well as bone conduction oral appliances and scalp-worn devices. The recommendation for BCHD should be determined by a qualified otolaryngology-head and neck surgeon. These devices are approved by the Food and Drug Administration (FDA) for these indications, and their use should adhere to the restrictions and guidelines specified by the appropriate governing agency, such as the FDA in the United States and the respective regulatory agencies in countries other than the United States."

U.S. Preventive Services Task Force Recommendations
Not applicable

Ongoing and Unpublished Clinical Trials
Some currently ongoing and unpublished trials that might influence this review are listed in Table 4.

Table 4. Summary of Key Trials

NCT No. Trial Name Planned Enrollment Completion Date
Ongoing      
NCT03533686 Comparative Study of Non-invasive Adhear Bone Conduction System to Traditional Bone Conduction Hearing Devices 90 Sep 2022 (Status = Recruiting)
Unpublished      
NCT01858246 A Randomised Controlled Trial Comparing Bone Anchored Hearing Aid With Bonebridge 60 Dec 2017
(Completed, last updated 2018)
NCT02022085a Post-market Clinical Follow-up of a Magnetic Bone Conduction Implant (Cochlear Baha Attract System) 2 Nov 2017
(Completed, last updated 2018)


NCT: national clinical trial.
a Denotes industry-sponsored or cosponsored trial.

References: 

  1. Janssen RM, Hong P, Chadha NK. Bilateral bone-anchored hearing aids for bilateral permanent conductive hearing loss: a systematic review. Otolaryngol Head Neck Surg. Sep 2012; 147(3): 412-22. PMID 22714424
  2. Bosman AJ, Snik AF, van der Pouw CT, et al. Audiometric evaluation of bilaterally fitted bone-anchored hearing aids. Audiology. May-Jun 2001; 40(3): 158-67. PMID 11465298
  3. Priwin C, Stenfelt S, Granstrom G, et al. Bilateral bone-anchored hearing aids (BAHAs): an audiometric evaluation. Laryngoscope. Jan 2004; 114(1): 77-84. PMID 14709999
  4. Snik AF, Mylanus EA, Proops DW, et al. Consensus statements on the BAHA system: where do we stand at present?. Ann Otol Rhinol Laryngol Suppl. Dec 2005; 195: 2-12. PMID 16619473
  5. Dun CA, de Wolf MJ, Mylanus EA, et al. Bilateral bone-anchored hearing aid application in children: the Nijmegen experience from 1996 to 2008. Otol Neurotol. Jun 2010; 31(4): 615-23. PMID 20393374
  6. Ho EC, Monksfield P, Egan E, et al. Bilateral Bone-anchored Hearing Aid: impact on quality of life measured with the Glasgow Benefit Inventory. Otol Neurotol. Oct 2009; 30(7): 891-6. PMID 19692937
  7. Briggs R, Van Hasselt A, Luntz M, et al. Clinical performance of a new magnetic bone conduction hearing implant system: results from a prospective, multicenter, clinical investigation. Otol Neurotol. Jun 2015; 36(5): 834-41. PMID 25634465
  8. Denoyelle F, Coudert C, Thierry B, et al. Hearing rehabilitation with the closed skin bone-anchored implant Sophono Alpha1: results of a prospective study in 15 children with ear atresia. Int J Pediatr Otorhinolaryngol. Mar 2015; 79(3): 382-7. PMID 25617189
  9. Hol MK, Nelissen RC, Agterberg MJ, et al. Comparison between a new implantable transcutaneous bone conductor and percutaneous bone-conduction hearing implant. Otol Neurotol. Aug 2013; 34(6): 1071-5. PMID 23598702
  10. Nelissen RC, Agterberg MJ, Hol MK, et al. Three-year experience with the Sophono in children with congenital conductive unilateral hearing loss: tolerability, audiometry, and sound localization compared to a bone-anchored hearing aid. Eur Arch Otorhinolaryngol. Oct 2016; 273(10): 3149-56. PMID 26924741
  11. Iseri M, Orhan KS, Tuncer U, et al. Transcutaneous Bone-anchored Hearing Aids Versus Percutaneous Ones: Multicenter Comparative Clinical Study. Otol Neurotol. Jun 2015; 36(5): 849-53. PMID 25730451
  12. Gerdes T, Salcher RB, Schwab B, et al. Comparison of Audiological Results Between a Transcutaneous and a Percutaneous Bone Conduction Instrument in Conductive Hearing Loss. Otol Neurotol. Jul 2016; 37(6): 685-91. PMID 27093021
  13. Dimitriadis PA, Farr MR, Allam A, et al. Three year experience with the cochlear BAHA attract implant: a systematic review of the literature. BMC Ear Nose Throat Disord. 2016; 16: 12. PMID 27733813
  14. Reddy-Kolanu R, Gan R, Marshall AH. A case series of a magnetic bone conduction hearing implant. Ann R Coll Surg Engl. Nov 2016; 98(8): 552-553. PMID 27490984
  15. Siegert R. Partially implantable bone conduction hearing aids without a percutaneous abutment (Otomag): technique and preliminary clinical results. Adv Otorhinolaryngol. 2011; 71: 41-46. PMID 21389703
  16. Powell HR, Rolfe AM, Birman CS. A Comparative Study of Audiologic Outcomes for Two Transcutaneous Bone-Anchored Hearing Devices. Otol Neurotol. Sep 2015; 36(9): 1525-31. PMID 26375976
  17. O'Niel MB, Runge CL, Friedland DR, et al. Patient Outcomes in Magnet-Based Implantable Auditory Assist Devices. JAMA Otolaryngol Head Neck Surg. Jun 2014; 140(6): 513-20. PMID 24763485
  18. Centric A, Chennupati SK. Abutment-free bone-anchored hearing devices in children: initial results and experience. Int J Pediatr Otorhinolaryngol. May 2014; 78(5): 875-8. PMID 24612554
  19. Baker S, Centric A, Chennupati SK. Innovation in abutment-free bone-anchored hearing devices in children: Updated results and experience. Int J Pediatr Otorhinolaryngol. Oct 2015; 79(10): 1667-72. PMID 26279245
  20. Marsella P, Scorpecci A, Vallarino MV, et al. Sophono in Pediatric Patients: The Experience of an Italian Tertiary Care Center. Otolaryngol Head Neck Surg. Aug 2014; 151(2): 328-32. PMID 24714216
  21. Magliulo G, Turchetta R, Iannella G, et al. Sophono Alpha System and subtotal petrosectomy with external auditory canal blind sac closure. Eur Arch Otorhinolaryngol. Sep 2015; 272(9): 2183-90. PMID 24908070
  22. Bravo-Torres S, Der-Mussa C, Fuentes-Lopez E. Active transcutaneous bone conduction implant: audiological results in paediatric patients with bilateral microtia associated with external auditory canal atresia. Int J Audiol. Jan 2018; 57(1): 53-60. PMID 28857620
  23. Schmerber S, Deguine O, Marx M, et al. Safety and effectiveness of the Bonebridge transcutaneous active direct-drive bone-conduction hearing implant at 1-year device use. Eur Arch Otorhinolaryngol. Apr 2017; 274(4): 1835-1851. PMID 27475796
  24. Rahne T, Seiwerth I, Gotze G, et al. Functional results after Bonebridge implantation in adults and children with conductive and mixed hearing loss. Eur Arch Otorhinolaryngol. Nov 2015; 272(11): 3263-9. PMID 25425039
  25. Laske RD, Roosli C, Pfiffner F, et al. Functional Results and Subjective Benefit of a Transcutaneous Bone Conduction Device in Patients With Single-Sided Deafness. Otol Neurotol. Aug 2015; 36(7): 1151-6. PMID 26111077
  26. Riss D, Arnoldner C, Baumgartner WD, et al. Indication criteria and outcomes with the Bonebridge transcutaneous bone-conduction implant. Laryngoscope. Dec 2014; 124(12): 2802-6. PMID 25142577
  27. Manrique M, Sanhueza I, Manrique R, et al. A new bone conduction implant: surgical technique and results. Otol Neurotol. Feb 2014; 35(2): 216-20. PMID 24448280
  28. Ihler F, Volbers L, Blum J, et al. Preliminary functional results and quality of life after implantation of a new bone conduction hearing device in patients with conductive and mixed hearing loss. Otol Neurotol. Feb 2014; 35(2): 211-5. PMID 24448279
  29. Desmet J, Wouters K, De Bodt M, et al. Long-term subjective benefit with a bone conduction implant sound processor in 44 patients with single-sided deafness. Otol Neurotol. Jul 2014; 35(6): 1017-25. PMID 24751733
  30. Iseri M, Orhan KS, Kara A, et al. A new transcutaneous bone anchored hearing device - the Baha(R) Attract System: the first experience in Turkey. Kulak Burun Bogaz Ihtis Derg. Mar-Apr 2014; 24(2): 59-64. PMID 24835899
  31. Peters JP, Smit AL, Stegeman I, et al. Review: Bone conduction devices and contralateral routing of sound systems in single-sided deafness. Laryngoscope. Jan 2015; 125(1): 218-26. PMID 25124297
  32. Baguley DM, Bird J, Humphriss RL, et al. The evidence base for the application of contralateral bone anchored hearing aids in acquired unilateral sensorineural hearing loss in adults. Clin Otolaryngol. Feb 2006; 31(1): 6-14. PMID 16441794
  33. den Besten CA, Monksfield P, Bosman A, et al. Audiological and clinical outcomes of a transcutaneous bone conduction hearing implant: Six-month results from a multicentre study. Clin Otolaryngol. Mar 2019; 44(2): 144-157. PMID 30358920
  34. Leterme G, Bernardeschi D, Bensemman A, et al. Contralateral routing of signal hearing aid versus transcutaneous bone conduction in single-sided deafness. Audiol Neurootol. 2015; 20(4): 251-60. PMID 26021779
  35. Snapp HA, Holt FD, Liu X, et al. Comparison of Speech-in-Noise and Localization Benefits in Unilateral Hearing Loss Subjects Using Contralateral Routing of Signal Hearing Aids or Bone-Anchored Implants. Otol Neurotol. Jan 2017; 38(1): 11-18. PMID 27846038
  36. Zeitler DM, Snapp HA, Telischi FF, et al. Bone-anchored implantation for single-sided deafness in patients with less than profound hearing loss. Otolaryngol Head Neck Surg. Jul 2012; 147(1): 105-11. PMID 22368043
  37. Pai I, Kelleher C, Nunn T, et al. Outcome of bone-anchored hearing aids for single-sided deafness: a prospective study. Acta Otolaryngol. Jul 2012; 132(7): 751-5. PMID 22497318
  38. Saroul N, Nicolas S, Akkari M, et al. Long-term benefit and sound localization in patients with single-sided deafness rehabilitated with an osseointegrated bone-conduction device. Otol Neurotol. Jan 2013; 34(1): 111-4. PMID 23202156
  39. Lin LM, Bowditch S, Anderson MJ, et al. Amplification in the rehabilitation of unilateral deafness: speech in noise and directional hearing effects with bone-anchored hearing and contralateral routing of signal amplification. Otol Neurotol. Feb 2006; 27(2): 172-82. PMID 16436986
  40. Kunst SJ, Leijendeckers JM, Mylanus EA, et al. Bone-anchored hearing aid system application for unilateral congenital conductive hearing impairment: audiometric results. Otol Neurotol. Jan 2008; 29(1): 2-7. PMID 18199951
  41. Kunst SJ, Hol MK, Mylanus EA, et al. Subjective benefit after BAHA system application in patients with congenital unilateral conductive hearing impairment. Otol Neurotol. Apr 2008; 29(3): 353-58. PMID 18494142
  42. Gluth MB, Eager KM, Eikelboom RH, et al. Long-term benefit perception, complications, and device malfunction rate of bone-anchored hearing aid implantation for profound unilateral sensorineural hearing loss. Otol Neurotol. Dec 2010; 31(9): 1427-34. PMID 20729779
  43. Faber HT, Nelissen RC, Kramer SE, et al. Bone-anchored hearing implants in single-sided deafness patients: Long-term use and satisfaction by gender. Laryngoscope. Dec 2015; 125(12): 2790-5. PMID 26152833
  44. Monini S, Musy I, Filippi C, et al. Bone conductive implants in single-sided deafness. Acta Otolaryngol. Apr 2015; 135(4): 381-8. PMID 25720582
  45. Amonoo-Kuofi K, Kelly A, Neeff M, et al. Experience of bone-anchored hearing aid implantation in children younger than 5 years of age. Int J Pediatr Otorhinolaryngol. Apr 2015; 79(4): 474-80. PMID 25680294
  46. Marsella P, Scorpecci A, Pacifico C, et al. Pediatric BAHA in Italy: the "Bambino Gesu" Children's Hospital's experience. Eur Arch Otorhinolaryngol. Feb 2012; 269(2): 467-74. PMID 21739094
  47. Davids T, Gordon KA, Clutton D, et al. Bone-anchored hearing aids in infants and children younger than 5 years. Arch Otolaryngol Head Neck Surg. Jan 2007; 133(1): 51-5. PMID 17224524
  48. McDermott AL, Williams J, Kuo MJ, et al. The role of bone anchored hearing aids in children with Down syndrome. Int J Pediatr Otorhinolaryngol. Jun 2008; 72(6): 751-7. PMID 18433885
  49. Schwab B, Wimmer W, Severens JL, et al. Adverse events associated with bone-conduction and middle-ear implants: a systematic review. Eur Arch Otorhinolaryngol. Feb 2020; 277(2): 423-438. PMID 31749056
  50. Verheij E, Bezdjian A, Grolman W, et al. A Systematic Review on Complications of Tissue Preservation Surgical Techniques in Percutaneous Bone Conduction Hearing Devices. Otol Neurotol. Aug 2016; 37(7): 829-37. PMID 27273402
  51. Kiringoda R, Lustig LR. A meta-analysis of the complications associated with osseointegrated hearing aids. Otol Neurotol. Jul 2013; 34(5): 790-4. PMID 23739555
  52. Dun CA, Faber HT, de Wolf MJ, et al. Assessment of more than 1,000 implanted percutaneous bone conduction devices: skin reactions and implant survival. Otol Neurotol. Feb 2012; 33(2): 192-8. PMID 22246385
  53. Hobson JC, Roper AJ, Andrew R, et al. Complications of bone-anchored hearing aid implantation. J Laryngol Otol. Feb 2010; 124(2): 132-6. PMID 19968889
  54. Wallberg E, Granstrom G, Tjellstrom A, et al. Implant survival rate in bone-anchored hearing aid users: long-term results. J Laryngol Otol. Nov 2011; 125(11): 1131-5. PMID 21774847
  55. Kraai T, Brown C, Neeff M, et al. Complications of bone-anchored hearing aids in pediatric patients. Int J Pediatr Otorhinolaryngol. Jun 2011; 75(6): 749-53. PMID 21470698
  56. Allis TJ, Owen BD, Chen B, et al. Longer length Baha abutments decrease wound complications and revision surgery. Laryngoscope. Apr 2014; 124(4): 989-92. PMID 24114744
  57. Calvo Bodnia N, Foghsgaard S, Nue Moller M, et al. Long-term results of 185 consecutive osseointegrated hearing device implantations: a comparison among children, adults, and elderly. Otol Neurotol. Dec 2014; 35(10): e301-6. PMID 25122598
  58. Rebol J. Soft tissue reactions in patients with bone anchored hearing aids. Ir J Med Sci. Jun 2015; 184(2): 487-91. PMID 24913737
  59. Larsson A, Tjellstrom A, Stalfors J. Implant losses for the bone-anchored hearing devices are more frequent in some patients. Otol Neurotol. Feb 2015; 36(2): 336-40. PMID 24809279
  60. den Besten CA, Nelissen RC, Peer PG, et al. A Retrospective Cohort Study on the Influence of Comorbidity on Soft Tissue Reactions, Revision Surgery, and Implant Loss in Bone-anchored Hearing Implants. Otol Neurotol. Jun 2015; 36(5): 812-8. PMID 25811351
  61. Mohamad S, Khan I, Hey SY, et al. A systematic review on skin complications of bone-anchored hearing aids in relation to surgical techniques. Eur Arch Otorhinolaryngol. Mar 2016; 273(3): 559-65. PMID 25503356
  62. Fontaine N, Hemar P, Schultz P, et al. BAHA implant: implantation technique and complications. Eur Ann Otorhinolaryngol Head Neck Dis. Feb 2014; 131(1): 69-74. PMID 23835074
  63. Hultcrantz M, Lanis A. A five-year follow-up on the osseointegration of bone-anchored hearing device implantation without tissue reduction. Otol Neurotol. Sep 2014; 35(8): 1480-5. PMID 24770406
  64. Nelissen RC, Stalfors J, de Wolf MJ, et al. Long-term stability, survival, and tolerability of a novel osseointegrated implant for bone conduction hearing: 3-year data from a multicenter, randomized, controlled, clinical investigation. Otol Neurotol. Sep 2014; 35(8): 1486-91. PMID 25080037
  65. Singam S, Williams R, Saxby C, et al. Percutaneous bone-anchored hearing implant surgery without soft-tissue reduction: up to 42 months of follow-up. Otol Neurotol. Oct 2014; 35(9): 1596-600. PMID 25076228
  66. Roplekar R, Lim A, Hussain SS. Has the use of the linear incision reduced skin complications in bone-anchored hearing aid implantation?. J Laryngol Otol. Jun 2016; 130(6): 541-4. PMID 27160014
  67. American Academy of Otolaryngology-Head and Neck Surgery. Position Statement: Bone Conduction Hearing Devices. Position Statements 2016; http://www.entnet.org/content/position-statement-bone-conduction-hearing- devices. Accessed January 31, 2022.
  68. Centers for Medicare & Medicaid Services. Medicare Policy Benefit Manual. Chapter 16 - General Exclusions from Coverage (Rev. 198). 2014; Rev. 189. http://www.cms.gov/manuals/Downloads/bp102c16.pdf. Accessed January 31, 2022.
  69. Centers for Medicare & Medicaid Services. Fact sheets: CMS Updates Policies and Payment Rates for End- Stage Renal Disease Facilities for CY 2015 and Implementation of Competitive Bidding-Based Prices for Durable Medical Equipment, Prosthetics, Orthotics, and Supplies. 2014; https://www.cms.gov/newsroom/fact-sheets/cms-updates-policies-and-payment-rates-end-stage-renal-disease-facilities-cy-2015-and-implementation. Accessed January 31, 2022.

Coding Section

Codes

Number

Description

CPT

69710

Implantation or replacement of electromagnetic bone conduction hearing device in temporal bone

 

69711

Removal or repair of electromagnetic bone conduction hearing device in temporal bone

 

69714

Implantation, osseointegrated implant, temporal bone, with percutaneous attachment to external speech processor/cochlear stimulator; without mastoidectomy

 

69715 (Effective 01/01/2022 code will be DELETED)

Implantation, osseointegrated implant, temporal bone, with percutaneous attachment to external speech processor/cochlear stimulator; with mastoidectomy

 

69716 (code effective 01/01/2022) 

Osseointegrated implant insertion with magnetic transcutaneous attachment to a speech processor 

 

69717

Replacement (including removal of existing device), osseointegrated implant, temporal bone, with percutaneous attachment to external speech processor/cochlear stimulator; without mastoidectomy

 

69718 (Effective 01/01/2022 code will be DELETED)

Replacement (including removal of existing device), osseointegrated implant, temporal bone, with percutaneous attachment to external speech processor/cochlear stimulator; with mastoidectomy

 

69719 (code effective 01/01/2022)

Osseointegrated implant revision or replacement with magnetic transcutaneous attachment to a speech processor 

 

69726 (code effective 01/01/2022)

Osseointegrated implant removal with percutaneous attachment to a speech processor 

 

69727 (code effective 01/01/2022)

Osseointegrated implant removal with magnetic transcutaneous attachment to a speech processor 

 

69799

Unlisted procedure, middle ear (use for BAHA Attract"

HCPCS

L8625

External recharging system for battery for use with cochlear implant or auditory osseointegrated device, replacement only, each

 

L8690

Auditory osseointegrated device, includes all internal and external components

 

L8691

Auditory osseointegrated device, external sound processor, replacement

 

L8693

Auditory osseointegrated device abutment, any length, replacement only

 

L8694

Auditory osseointegrated device, transducer/actuator, replacement only, each

ICD-10-CM

H60.00-H60.93

Otitis externa

 

H61.301-H61.399

Acquired stenosis of external ear canal code range

 

H65.20-H65.499; H66.10-H66.3x9

Chronic otitis media code range

 

H90.0-H90.8

Conductive and sensorineural hearing loss code range

 

Q16.0-Q16.9

Congenital malformations of ear causing impairment of hearing, code range

ICD-10-PCS

 

ICD-10-PCS codes are only used for inpatient services.

 

09HD05Z, 09HE05Z09HD06Z, 09HE06Z, 09PD0SZ, 09PE0SZ, 09BB0ZZ, 09BC0ZZ, 09HD06Z, 09HE06Z, 09PD0SZ, 09PE0SZ

Surgical, ear, nose & sinus, insertion, inner ear, open, hearing device, bone conduction, code for left or right

 

0NH50SZ, 0NH53SZ, 0NH54SZ, 0NH60SZ, 0NH63SZ, 0NH64SZ

Surgical, head & facial bones, insertion, temporal bone, hearing device, code by left or right, and approach (open, percutaneous, percutaneous endoscopic)

Type of Service

Surgery

 

Place of Service

Outpatient

 

Procedure and diagnosis codes on Medical Policy documents are included only as a general reference tool for each policy. They may not be all-inclusive. 

Index
Audiant™ Bone Conductor
Bone Anchored Hearing Aids, Semi-implantable
Hearing Aids, Bone Conduction, Implantable

This medical policy was developed through consideration of peer-reviewed medical literature generally recognized by the relevant medical community, U.S. FDA approval status, nationally accepted standards of medical practice and accepted standards of medical practice in this community, Blue Cross Blue Shield Association technology assessment program (TEC) and other nonaffiliated technology evaluation centers, reference to federal regulations, other plan medical policies, and accredited national guidelines.

"Current Procedural Terminology © American Medical Association. All Rights Reserved" 

History From 2013 Forward     

12/15/2022 Annual review, no change to policy. Updating rationale and references.

12/01/2021 

Annual review, no change to policy intent. Updating rationale and references. 

11/29/2021

Updating policy with 2022 coding. Adding code 69716, 69719, 69726 and 69727. On 01/01/2022 the following codes will be DELETED 69715 and 69718.

12/01/2020

Annual review, policy intent unchanged. Updating regulatory status, guidelines, coding, rationale and references. 

12/10/2019 

Annual review, no change to policy intent. Updating background, regulatory status, guidelines, rationale and references. 

12/13/2018 

Annual review, no change to policy intent. Updating rationale and references. 

12/14/2017

Annual review, no change to policy intent. Updating regulatory status, rationale and references. 

12/01/2016 

Annual review, updating policy verbiage to remove investigational status from partially implantable devices. Also updating background, description, regulatory status, rationale and references. 

11/12/2015 

Annual review, no change to policy intent. Updated background, description, regulatory status, rationale and references. 

12/10/2014 

Annual review, no change to policy intent. Added key words and coding. Updated policy guidelines, regulatory status, rationale and references.

12/2/2013

Added verbiage and references to indicate that intra oral bone conduction hearing aids are investigational.